BOSTON SCIENTIFIC CORPORATION SYNERGY; BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Model Number 10603
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: used the first date of the month of the aware date as no event date was provided.Device evaluated by mfr.: synergy ii us mr 3.00 x 20mm stent delivery system was returned for analysis.A visual examination of the stent found evidence of damage.Struts along the proximal end were pulled distally.The undamaged crimped stent outer diameter was measured and the result is within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found no issues.A visual examination of the outer and inner lumen and mid-shaft section found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
 
Event or Problem Description
Reportable based on device analysis completed on 01apr2021.It was reported that crossing difficulties were encountered.A 3.00 x 20 synergy ii drug-eluting stent was advanced but failed to cross the lesion.The device was completely removed and the procedure was completed with another of the same device.There were no patient complications nor injuries reported and the patient's status was stable.However, returned device analysis revealed a stent damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11704477
Report Number2134265-2021-05150
Device Sequence Number17226098
Product Code NIQ
UDI-Device Identifier08714729840176
UDI-Public08714729840176
Combination Product (Y/N)Y
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/24/2022
Device Model Number10603
Device Catalogue Number10603
Device Lot Number0025958145
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/01/2021
Initial Report FDA Received Date04/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-