C.R. BARD, INC. (BASD) -3006260740 POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Model Number |
1829570 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Protrusion/Extrusion (2979)
|
| Health Effect - Clinical Code |
Extravasation (1842)
|
| Date of Event |
03/26/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 03/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that post port device implant, the septum of the device popped off when being power injected.The patient experienced extravasation of contrast under the skin and the port was removed and replaced with another device.The current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: a lot history review, a device history record review and complete manufacturing review could not be conducted for the investigation as the lot number is unknown.Investigation summary: the device was not returned for evaluation.The investigation is inconclusive for the reported extravasation and material protrusion issue.Furthermore, the clinical condition alleged in the complaint cannot be confirmed.A definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that post port device implant, the septum of the device popped off when being power injected.The patient experienced extravasation of contrast under the skin and the port was removed and replaced with another device.The current status of the patient is unknown.
|
| |
|
Search Alerts/Recalls
|
|
|