| Model Number |
VENUM14160 |
| Medical Device Problem Code |
Activation Failure (3270)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.(expiry date: 10/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that during a stent deployment procedure, the distal end of the stent allegedly failed to expand.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the lot history record of this lot was reviewed, with special attention to the manufacturing and inspection of this product.And the product was found to have met the specification prior to shipment.Investigation summary: the delivery system was returned for evaluation.The condition of the returned delivery system confirmed, an expansion issue.The stent brake of the inner catheter, which had been under the stent, during deployment was found with heavy imprints, and superficial damage.Which was considered, an indication that the stent adhered to the stent brake after deployment.Which leads to a confirmed, result for expansion issue.A definite root cause for the reported event could not be determined.Labeling review: a review of the relevant instruction for use for this product was conducted.The instruction for use was found to address potential complications and adverse events such as incorrect positioning of the stent requiring further stenting or surgery, intimal injury, and malposition.H10: d4 (expiry date: 10/2022).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported, that during a stent deployment procedure.The distal end of the stent allegedly, failed to expand.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|