• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AUTOCAT2 WAVE JAPANESE; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AUTOCAT2 WAVE JAPANESE; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000325
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 03/26/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn# (b)(4).
 
Event or Problem Description
There was no patient involvement.It was reported that during operational checking of the intra-aortic balloon pump (iabp) (battery- operated) at the distributor's site, after about 60 minutes, a "large helium leak" alarm and "high base line" alarm occurred.
 
Additional Manufacturer Narrative
(b)(4).Teleflex received the device for investigation.The reported complaint of "high baseline and helium loss alarm" is not confirmed.The returned m-force motor driver assembly passed visual and functional test specifications.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the lot number/serial number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
 
Event or Problem Description
There was no patient involvement.It was reported that during operational checking of the intra-aortic balloon pump (iabp) (battery- operated) at the distributor's site, after about 60 minutes, a "large helium leak" alarm and "high base line" alarm occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT2 WAVE JAPANESE
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key11706599
Report Number3010532612-2021-00099
Device Sequence Number11736151
Product Code DSP
UDI-Device Identifier30801902043420
UDI-Public30801902043420
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K060309
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 03/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN000325
Device Catalogue NumberIAP-0500J
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2021
Initial Date Received by Manufacturer 03/31/2021
Supplement Date Received by Manufacturer06/14/2021
Initial Report FDA Received Date04/22/2021
Supplement Report FDA Received Date06/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-