MENTOR TEXAS MENTOR SILTEX CONTOUR PROFILE MODERATE; PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
|
Back to Search Results |
|
| Model Number |
3542911 |
| Medical Device Problem Code |
Migration (4003)
|
| Health Effect - Clinical Code |
Wrinkling (2613)
|
| Date of Event |
03/10/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: cutaneous contour deformity and device migration.Manufacturer¿s reference number: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a (b)(6) year-old caucasian female patient underwent a breast reconstruction primary with a mentor siltex contour profile moderate 225cc and suffered from a cutaneous contour deformity on the right side.The patient experienced a concern for position and width of implant.As a result, the patient had undergone removal and replacement with mentor smooth round high profile 210cc on (b)(6) 2021.
|
| |
|
Additional Manufacturer Narrative
|
|
On (b)(6) 2021, mentor conducted a visual inspection and leak testing of the returned device.During the visual evaluation, no apparent damage or visual anomalies were observed on the con pro 225/250cc, 225cc returned device.Leak testing was performed, in accordance with mentor procedures, and no leak sites were detected during the analysis.Implant displacement/migration may occur from improper implant sizing and/or placement, i.E., when the implant is too large or the pocket too small or when there has been inadequate preoperative assessment of stresses causing movement of the prosthesis.Implant displacement/migration is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor¿s quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 5/4/2021, the mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
|
| |
|