• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AVANOS MEDICAL INC. MIC PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PEG KIT WITH ENFIT CONNECTORS - 14 FR; DH EF PEG INITIAL PLACEMENT PRODUCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AVANOS MEDICAL INC. MIC PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PEG KIT WITH ENFIT CONNECTORS - 14 FR; DH EF PEG INITIAL PLACEMENT PRODUCTS Back to Search Results
Model Number 8160-14
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Codes Necrosis (1971); Peritonitis (2252); Fasciitis (2375); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 03/18/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The actual complaint product was not returned for evaluation.A review of the device history record is not possible as no lot number was provided.Root cause could not be determined.All information reasonably known as of 22-apr-2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).
 
Event or Problem Description
It was reported "the inner disc [of the peg device] was within the stomach, but the external bumper had slipped, permitting the stomach to pull away thereby producing the peritonitis, compartment syndrome and now necrotizing fasciitis." additional information received 08-apr-2021 stated the patient underwent an emergency laparotomy with replacement of the peg tube with a gastrostomy tube.The patient remained critically ill in the cardiac intensive care unit (icu) with necrotizing fasciitis and had undergone multiple surgical procedures.
 
Event or Problem Description
Additional information was received on 27-apr-2021 from fda medwatch report # 5100693 and the following was noted: a female underwent uncomplicated peg placement with a 14 fr avanos brand mic percutaneous endoscopic gastrostomy the patient had migration of the tube, allowing formula leakage, peritonitis, sepsis and now necrotizing fasciitis (noted via (b)(4)) we had a similar event 1 week later with the same size and brand tube without other explanatory factors on review of data, ((b)(4)) this is the 3rd tube migration with this size and brand in a 12 month period at our institution.
 
Additional Manufacturer Narrative
Correction: b3, d1, d4 additional information: b5, d4, g2 and h4 the device history record for the reported lot number, 30062781, in this complaint was reviewed and the material was produced according to the manufacturing procedures and met the quality requirements.All information reasonably known as of 24-may-2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Event or Problem Description
Additional information received 02-sep-2021 stating follow-up with the physician noted the common cause in the case was that no pexy was done.The gastroenterologist works with a general surgeon on the procedure to perform the endoscopy and stoma creation/peg placement respectively; and importantly that pexies have never been part of our procedure.The patient was medically complicated with underlying health issues, cardiac, etc.The stoma came apart around 3-days post-operatively.The external disc never slipped but the physician felt the internal bumper was 'too floppy.'.
 
Additional Manufacturer Narrative
All information reasonably known as of 23-sep-2021 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical, inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical, inc.Avanos medical, inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the avanos medical, inc.Complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical, inc.Product is defective or caused serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIC PERCUTANEOUS ENDOSCOPIC GASTROSTOMY PEG KIT WITH ENFIT CONNECTORS - 14 FR
Common Device Name
DH EF PEG INITIAL PLACEMENT PRODUCTS
Manufacturer (Section D)
AVANOS MEDICAL INC.
5405 windward parkway
alpharetta GA 30004
MDR Report Key11711212
Report Number3006646024-2021-00004
Device Sequence Number17003665
Product Code KNT
UDI-Device Identifier00350770446110
UDI-Public00350770446110
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K924065
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model Number8160-14
Device Catalogue NumberN/A
Device Lot Number30062781
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/01/2021
Supplement Date Received by Manufacturer04/27/2021
09/02/2021
Initial Report FDA Received Date04/22/2021
Supplement Report FDA Received Date05/25/2021
09/24/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening; Required Intervention;
Patient Age2 YR
Patient Weight10
-
-