| Model Number |
SSLP-400 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
10/15/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side capsular contracture baker grade iv.Device has been explanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported a right side capsular contracture baker grade iv.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: the weight the device within specification, flat creases, and a curved opening on radius.A microscopic analysis was performed which identified: two striated openings on radius.Based on the device analysis the final assessment is: two striated openings on radius assessed as surgical damage consist in the use of some surgical tool.
|
| |
|