• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALLERGAN (COSTA RICA) INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Model Number SSLP-400
Medical Device Problem Code Device Appears to Trigger Rejection (1524)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 10/15/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iv.
 
Event or Problem Description
Healthcare professional reported a right side capsular contracture baker grade iv.Device has been explanted.
 
Event or Problem Description
Healthcare professional reported a right side capsular contracture baker grade iv.Device has been explanted.
 
Additional Manufacturer Narrative
Device evaluation: visual analysis of the returned device identified: the weight the device within specification, flat creases, and a curved opening on radius.A microscopic analysis was performed which identified: two striated openings on radius.Based on the device analysis the final assessment is: two striated openings on radius assessed as surgical damage consist in the use of some surgical tool.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INSPIRA SMOOTH SILICONE GEL FILLED BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
ALLERGAN (COSTA RICA)
900 parkway global park
zona franca
MDR Report Key11711283
Report Number9617229-2021-06263
Device Sequence Number3894565
Product Code FTR
UDI-Device Identifier10888628032774
UDI-Public10888628032774
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P020056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/15/2025
Device Model NumberSSLP-400
Device Catalogue NumberSSLP-400
Device Lot Number3445431
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2021
Initial Date Received by Manufacturer 03/24/2021
Supplement Date Received by Manufacturer05/12/2021
Initial Report FDA Received Date04/22/2021
Supplement Report FDA Received Date06/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age42 YR
Patient Weight119
-
-