| Model Number |
CTL4.5-040-S |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation as it remains in-situ.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
Information was received that a revision procedure was performed in (b)(6) 2021 to address loose screws.As per the reporter, approximately two and a half weeks post-implantation, it was discovered that the screws in the proximal bone segment had loosened and disengaged the plate.No patient adverse event was reported.During the revision, the screws were re-tightened and the screw holes were back-filled with cement to prevent future screw back-out.Post revision, the patient has successfully resumed lengthening.Report 3 of 3.
|
| |
|