| Catalog Number |
4C8723 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that 3 clearlink system y-type blood/solution set were defective.The defect was further described as leaked at the hand pump.This issue was identified during patient infusion.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
One (1) actual device was received for evaluation.Visual inspection was performed using the naked eye did not identify any abnormalities that could have contributed to the reported condition.All components were correctly placed and according to specifications.Functional testing was performed including clear passage and pressure testing which revealed a leak in the hand pump between the assembly of the top cap and barrel tubing.The reported condition was verified.The cause of the condition was due to improper bonding during manufacturing process.The other two (2) samples were not received and therefore, could not be evaluated.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|