MEDOS INTERNATIONAL SàRL CH 5.5 EXP VERSE SCREW 6.0 X 35; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 199721635
Medical Device Problem Codes Difficult to Advance (2920); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional device product codes: kwp;kwq;mnh;mni;osh.Complainant part is not expected to be returned for manufacturer review/investigation.The investigation could not be completed; no conclusion could be drawn, as no product was received.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that during a procedure on (b)(6) 2021, surgeon tried to insert the locking screws into the pedicle screws, but felt that they did not trap.After few attempts, surgeon noticed that the locking screws were disintegrating, creating metal filings that the team tried to fully recover.No further information provided.This report is for one (1) 5.5 exp verse screw 6.0 x 35.This is report 3 of 4 for complaint (b)(4).
 
Event or Problem Description
It was further reported that the procedure was completed by applying a 5.0x35 screw.Procedure was completed successfully with a delay of twenty (20) minutes.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h4, h6: a device history record (dhr) review was conducted: the dhr of product code: 199721635, lot : 276019 was electronically reviewed and no non-conformances were observed during the manufacturing process.The product was released on: 17.03.2020.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 EXP VERSE SCREW 6.0 X 35
Common Device Name
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key11712634
Report Number1526439-2021-00778
Device Sequence Number3193308
Product Code NKB
UDI-Device Identifier10705034442297
UDI-Public(01)10705034442297
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K142185
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number199721635
Device Catalogue Number199721635
Device Lot Number276019
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/29/2021
Supplement Date Received by Manufacturer04/26/2021
06/25/2021
Initial Report FDA Received Date04/23/2021
Supplement Report FDA Received Date05/21/2021
06/25/2021
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
5.5 EXP VERSE UNITIZED SET SCR.; UNKNOWN INSERTION INSTRUMENTS.; VERSE CORRECTION KEY.; VERSE CORRECTION KEY.
-
-