• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 CORAIL AMT SZ11 135 HIGH COL; CORAIL LINE EXTENSION IMPLANTS : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 CORAIL AMT SZ11 135 HIGH COL; CORAIL LINE EXTENSION IMPLANTS : HIP FEMORAL STEM Back to Search Results
Model Number L971111
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Fall (1848); Limb Fracture (4518)
Date of Event 04/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient fell, revised due to fracture.No surgical delay.Doi: unknown.Dor: (b)(6) 2021; left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the material, manufacturing, inspection or sterile processing that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CORAIL AMT SZ11 135 HIGH COL
Common Device Name
CORAIL LINE EXTENSION IMPLANTS : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key11712899
Report Number1818910-2021-08532
Device Sequence Number14394097
Product Code LZO
UDI-Device Identifier10603295455301
UDI-Public10603295455301
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K190344
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberL971111
Device Catalogue NumberL971111
Device Lot Number9367010
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/07/2021
Supplement Date Received by Manufacturer05/18/2021
Initial Report FDA Received Date04/23/2021
Supplement Report FDA Received Date05/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICUL/EZE BALL 28 +5 BR; SELF CENT HIP 45X28 GRY
Outcome Attributed to Adverse Event Required Intervention;
Patient Age85 YR
-
-