GYRUS ACMI, INC MULTIDEBRIDER POWER CONSOLE Back to Search Results
Model Number MDCONS100
Medical Device Problem Codes Failure to Power Up (1476); Output Problem (3005); Intermittent Energy Output (4025)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The returned device was evaluated.Inspection found that the reported issue was not confirmed.During testing, all output powers were observed to be within specifications, leakage test passed and no issues were found.In addition, burn in testing was performed and device was turned on and off for few times and the reported issue was not able to be duplicated.The unit passed all functional testing with no errors observed.Unrelated to the reported issue, minor scratches and dents were found on the housing unit and new version card edge board was noted on the device.Based on evaluation findings the reported issue was not able to be duplicated as no problem found on the device during testing.This report will be supplemented accordingly following investigations.
 
Event or Problem Description
A report of main console not functioning as intended, system is not booting up with bipolar mode working intermittently during preparation for use was reported.No patient involvement, no user injury reported due to the event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation.The following sections were updated: g3, g6, h2, h4, h6, and h10.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the legal manufacturer's investigation, the root cause has determined to be user error as the customer phenomenon could not be replicated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MULTIDEBRIDER POWER CONSOLE
Common Device Name
MULTIDEBRIDER POWER CONSOLE
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough MA 01772
MDR Report Key11713079
Report Number1037007-2021-00029
Device Sequence Number11720830
Product Code GEI
UDI-Device Identifier00821925027770
UDI-Public00821925027770
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K123429
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/11/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMDCONS100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/29/2021
Supplement Date Received by Manufacturer04/27/2021
Initial Report FDA Received Date04/23/2021
Supplement Report FDA Received Date05/11/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-