• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 S-ROM*HEAD FEMORAL COCR 28+0; S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 S-ROM*HEAD FEMORAL COCR 28+0; S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS Back to Search Results
Model Number 52-2028
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/07/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that this patient had her hip replaced 20+ years ago and had a revision done in 2005 by another dr.According to current dr.This patient recently began complaint of pain in her left hip.Through examination and testing (no labs or documents available with results) it was determined this patient has elevated chromium levels.Current dr.Scheduled the patient for a hip revision where he decided to do a hip ball exchange.The srom stem, srom sleeve, arthropor cup and liner were left in situ.Doi: unknown.Dor: (b)(6) 2021; affected side: left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot: a device manufacturing (mre) review will not be performed even when product/lot information is known.Per wi-3430 it has been determined that, for the mom platform and related allegations an mre is not required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM FEMORAL HEADS (11/13) : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key11713251
Report Number1818910-2021-08541
Device Sequence Number3766020
Product Code JDI
UDI-Device Identifier10603295170792
UDI-Public10603295170792
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K920317
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number52-2028
Device Catalogue Number522028
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/07/2021
Supplement Date Received by Manufacturer05/18/2021
Initial Report FDA Received Date04/23/2021
Supplement Report FDA Received Date05/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTHROPOR CUP; ARTHROPOR LINER; SROM SLEEVE; SROM STEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age63 YR
Patient Weight80
-
-