| Model Number |
66800042 |
| Medical Device Problem Code |
Pressure Problem (3012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a versajet exact assy, 45 degree x 8mm, the handpiece didn´t have pressure.The flow was not a one solid stream.The inspection was complete using a s+n back-up device.No case involved.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device intended for use in treatment was not returned for evaluation with all information provided we have not been able to establish a relationship between the device and the reported event or determine a root cause.A review of the manufacturing records found that there was no evidence that the product didn¿t meet specifications at the time of manufacture.A complaint history review found other related failures.Probable cause may be kinks, obstructions, or leaks.This investigation is now complete with no further action deemed necessary.Smith and nephew will continue to monitor for any adverse trends relating to this product range.
|
| |
|