• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INOGEN MINI ICD VR; IMPLANTABLE DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INOGEN MINI ICD VR; IMPLANTABLE DEVICE Back to Search Results
Model Number D011
Medical Device Problem Code Premature Discharge of Battery (1057)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.This device remains in service.No adverse patient effects were reported.
 
Event or Problem Description
It was reported that this implantable cardioverter defibrillator (icd) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.This device was explanted and returned for analysis.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.External visual inspection of the device revealed no anomalies.Review of the device memory confirmed that a low voltage alert was recorded.Although the device memory diagnostics demonstrated that the daily battery voltage measurements displayed an irregular pattern of discharge, the battery voltage level upon receipt in the laboratory was sufficient to ensure therapy availability/delivery while the device was implanted.The device case was then opened to facilitate analysis of the internal components.The battery was separated from the other device electronics and the overall current draw of the circuitry was measured.A normal current drain was observed within the circuitry.Collectively, the pattern of irregular daily battery voltage measurements in conjunction with normal power levels and device hybrid current draw is consistent with behavior of devices where a latent current leakage path has occurred within the battery itself, resulting in a partial depletion of the battery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN MINI ICD VR
Common Device Name
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key11713404
Report Number2124215-2021-05262
Device Sequence Number3527475
Product Code LWS
UDI-Device Identifier00802526533808
UDI-Public00802526533808
Combination Product (Y/N)N
Initial Reporter CountryTU
PMA/510(K) Number
P960040/S306
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date01/08/2018
Device Model NumberD011
Device Catalogue NumberD011
Device Lot Number103389
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2021
Initial Date Received by Manufacturer 02/16/2021
Supplement Date Received by Manufacturer05/03/2021
Initial Report FDA Received Date04/23/2021
Supplement Report FDA Received Date07/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-