| Catalog Number |
0004510129 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/24/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No sample has been received by manufacturing for evaluation.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.This is the seventh of seven reports for this reported event.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A nurse reported an "outbreak" of tass at their facility.In a follow up with the customer, it was reported that there were four patients who presented with tass following ocular surgery.There were two surgeons involved.All four patients associated with this report of tass had a history of glaucoma.The exact procedure performed on any one patient remains unconfirmed.This file represents one of the four patients.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned product was received at manufacturing.This complaint indication cannot be visually inspected.Returned product cannot confirm this complaint indication.Complaint trending was reviewed for the lot code provided.Two similar complaints were found.The complaint condition could not be confirmed.Batch records were reviewed and all testing results met specifications for this lot code at the time of release.Additionally, there were no deviations noted during batch record review.Review of the batch record indicates that the product meets specification.No manufacturing assignable root cause could be determined.There is no impact to the safety, quality or efficacy of the product.Based on the investigation, the lack of additional complaints for this lot, the returned sample evaluation and the acceptable batch record review, no further action is warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|