MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Model Number 5829502
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Codes Swelling/ Edema (4577); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/23/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 05/2022).
 
Event or Problem Description
It was reported that post port device implant, the device allegedly leaked.There was no reported patient injury.
 
Event or Problem Description
It was reported that post port device implant, the device allegedly leaked from the port access site and the patient experienced mild edema.The current status of the patient is unknown.
 
Additional Manufacturer Narrative
Manufacturing review: the device history record review was performed for the reported lot number and this lot meets all release criteria.The manufacturing review did not indicate any possible manufacturing issue that could be related to the reported event.Investigation summary: the device was not returned for evaluation.Therefore, the reported fluid leak issues cannot be confirmed.Based upon the available information a definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.Expiry date: 05/2022.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Additional Manufacturer Narrative
H10: manufacturing review: a complaint history review was performed.This is the thirty fourth complaint reported for this product/lot number combination.The device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for these reports.Investigation summary: the device was not returned for evaluation.Therefore, the reported fluid leak issues cannot be confirmed.Based upon the available information a definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 05/2022), g3.H11: h6 (result and conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
 
Event or Problem Description
It was reported that post port device implant, the device allegedly leaked.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F
Common Device Name
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key11714613
Report Number3006260740-2021-01495
Device Sequence Number3833126
Product Code LJT
UDI-Device Identifier00801741027406
UDI-Public(01)00801741027406
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K090512
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Risk Manager
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5829502
Device Catalogue Number5829502
Device Lot NumberREEV2356
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/29/2021
Supplement Date Received by Manufacturer05/11/2021
05/27/2021
Initial Report FDA Received Date04/23/2021
Supplement Report FDA Received Date05/18/2021
06/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age80 YR
Patient Weight53
-
-