C.R. BARD, INC. (BASD) -3006260740 POWERPORT DUO M.R.I. IMPLANTABLE PORT, CHRONOFLEX DUAL-LUMEN, 9.5F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
| Model Number |
5829502 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Codes |
Swelling/ Edema (4577); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 05/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that post port device implant, the device allegedly leaked.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that post port device implant, the device allegedly leaked from the port access site and the patient experienced mild edema.The current status of the patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Manufacturing review: the device history record review was performed for the reported lot number and this lot meets all release criteria.The manufacturing review did not indicate any possible manufacturing issue that could be related to the reported event.Investigation summary: the device was not returned for evaluation.Therefore, the reported fluid leak issues cannot be confirmed.Based upon the available information a definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.Expiry date: 05/2022.Section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a complaint history review was performed.This is the thirty fourth complaint reported for this product/lot number combination.The device history records were reviewed and there was nothing found to indicate there was a manufacturing related cause for these reports.Investigation summary: the device was not returned for evaluation.Therefore, the reported fluid leak issues cannot be confirmed.Based upon the available information a definitive root cause could not be determined.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 05/2022), g3.H11: h6 (result and conclusion).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that post port device implant, the device allegedly leaked.There was no reported patient injury.
|
| |
|
Search Alerts/Recalls
|
|
|