• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. CONQUEST PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. CONQUEST PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number CQ7594
Medical Device Problem Codes Deflation Problem (1149); Inflation Problem (1310)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/02/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: a complaint history review was performed.This is the first complaint reported to date for this product and lot, therefore a device history record review is not required.Investigation summary: one conquest pta device was returned for evaluation.The sample was bloody, no kinks noted to the catheter.Functional evaluation was performed, the device guidewire lumen was flushed.The balloon was inflated using an in-house presto inflation device.The balloon was able to hold pressure and shape.It was noted that the deflation was slow, the balloon did not fully deflate.The balloon was cut and under microscopic evaluation it was noted that the glue bullet was not seated properly and inflation port holes were noted to be collapsed.Therefore, the investigation is confirmed for the identified deflation issue as the glue bullet was not seated properly.The investigation is also confirmed for the inflation issue as collapsed port holes were noted.A definitive root cause could not be determined based upon available information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(expiry date: 09/2023).
 
Event or Problem Description
It was reported that during a procedure through the left arm, the balloon was allegedly unable to inflate after several attempts.The procedure was completed using another device.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONQUEST PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key11714693
Report Number2020394-2021-00905
Device Sequence Number3591088
Product Code DQY
UDI-Device Identifier00801741063190
UDI-Public(01)00801741063190
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K083657
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/31/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCQ7594
Device Catalogue NumberCQ7594
Device Lot NumberREEX0952
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/26/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 03/30/2021
Initial Report FDA Received Date04/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-