Model Number PCDG1 |
Device Problem
Peeled/Delaminated (1454)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/26/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).A photo of the product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent. if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
It was reported that a patient underwent a hernia repair procedure on (b)(6) 2021 and the mesh was used.It was found that the mesh was defective as the components separated visceral and parietal sides.It was reported that surgery was delayed by one hour.The procedure was successfully completed with a competitor device.There were no adverse patient consequences reported.
|
|
Manufacturer Narrative
|
Product complaint #: (b)(4).Date sent to the fda: 08/12/2021.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.The product was not returned to ethicon inc for evaluation.Upon visual inspection of the picture received, a foil was observed, however, the assignable cause of the reported condition cannot be determined in the picture.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint since the sample was not returned for analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4) date sent to the fda: 04/01/2022 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: d9, h3, h4, h6 h3 evaluation: one opened sample of product was returned for analysis.The mesh was noted to be as expected with no issues or anomalies.After further evaluation, the proceed mesh met the requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
|
Search Alerts/Recalls
|
|
|