• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. PROCEED*SURG MESH/MULTI LYR; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. PROCEED*SURG MESH/MULTI LYR; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number PCDG1
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/26/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A photo of the product upon which this medwatch is based has been received, however, the product evaluation is not yet complete.Any further information derived from the evaluation will be submitted in a supplemental 3500a form   attempts are being made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.  if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that a patient underwent a hernia repair procedure on (b)(6) 2021 and the mesh was used.It was found that the mesh was defective as the components separated visceral and parietal sides.It was reported that surgery was delayed by one hour.The procedure was successfully completed with a competitor device.There were no adverse patient consequences reported.
 
Manufacturer Narrative
Product complaint #: (b)(4).Date sent to the fda: 08/12/2021.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.The product was not returned to ethicon inc for evaluation.Upon visual inspection of the picture received, a foil was observed, however, the assignable cause of the reported condition cannot be determined in the picture.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.No conclusion could be reached as to what caused the reported complaint since the sample was not returned for analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4) date sent to the fda: 04/01/2022 this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information:  d9, h3, h4, h6 h3 evaluation: one opened sample of product was returned for analysis.The mesh was noted to be as expected with no issues or anomalies.After further evaluation, the proceed mesh met the requirements.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-CORNELIA
655 ethicon circle
cornelia GA 30531
Manufacturer Contact
elba bello
1000 route 202
raritan, NJ 08869
MDR Report Key11715128
MDR Text Key247060678
Report Number2210968-2021-03702
Device Sequence Number1
Product Code FTL
UDI-Device Identifier10705031047716
UDI-Public10705031047716
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K060713
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 04/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2020
Device Model NumberPCDG1
Device Catalogue NumberPCDG1
Device Lot NumberMKG236
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/27/2021
Initial Date FDA Received04/23/2021
Supplement Dates Manufacturer Received07/17/2021
03/03/2022
Supplement Dates FDA Received08/12/2021
04/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-