| Model Number |
10046799 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Laceration(s) (1946)
|
| Date of Event |
03/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reporting facility phone number and contact name were not provided to siemens.The event was reported by a siemens employee.Siemens performed technical evaluation of the reported event and system.The plexi-ring had aged and had become brittle, requiring a replacement which was performed.The system user manual cautions against performing maintenance and other tasks while the gantry is in motion.The root cause is use error.No device malfunction or general design were identified.
|
| |
|
Event or Problem Description
|
|
It was reported to siemens that the plexi-ring of the ct gantry cover has jumped out of its track due to a missing/broken plastic lock.The facility technician attempted to push the plexi-ring back into its original position but did not wait until the gantry stopped rotating.The technician's finger got caught in the rotating parts and a moderate injury to the digit was sustained.It was reported that both the skin and muscle of the digit were injured.The injury was treated by suturing the laceration.There was no permanent damage to the technician's finger.This incident is considered a work accident and no malfunction or general design issue were reported.The reported event occurred in (b)(6).
|
| |
|
Search Alerts/Recalls
|