ZIMMER BIOMET, INC. TPRLC 133 FP TYPE1 PPS SO 6.0; PROSTHETIC, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 -2021 -01227.0001825034 -2021 -01228.
 
Event or Problem Description
It was reported the distributor investigated circulated items and identified sterile packages damaged.No patients involved.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Event or Problem Description
No further event information available at the time of this report.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Evaluation of the returned product confirmed damage to the sterile packaging.Sterility has not been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the products when they left zimmer biomet was conforming to specifications.The root cause of the reported event is likely due to transit damage.Upon investigation, it has been determined that the damage to the sterile packaging meets the acceptable criteria and product is conforming to specifications.Event is no longer considered reportable, and initial report should be voided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 FP TYPE1 PPS SO 6.0
Common Device Name
PROSTHETIC, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11715920
Report Number0001825034-2021-01229
Device Sequence Number19894630
Product Code KWA
UDI-Device Identifier00880304498174
UDI-Public(01)00880304498174(17)270531(10)6061898
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K103755
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number51-100060
Device Lot Number6061898
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/05/2021
Supplement Date Received by Manufacturer06/03/2021
Initial Report FDA Received Date04/23/2021
Supplement Report FDA Received Date06/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-