Olympus recommended the device be returned for evaluation; however, the device has not been returned.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely the power supply and igniter unit failed due to aging because the subject device has been in operation for more than 10 years.This causes the main lamp to fail to illuminate, and it is assumed that the spare lamp is on intermittently.This mdr is being submitted retrospectively as part of a remediation effort related to recent system and process changes.A capa has been opened to manage the actions related to remediation of this issue and any required mdr reporting.
|