OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problems False Alarm (1013); Output Problem (3005)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation where service confirmed the customer's complaint was due to a faulty main circuit board.The device was repaired and returned to the customer.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the internal circuit malfunction is likely due to the failure of the pressure sensor mounted on the main board.The cause of the failure of the pressure sensor mounted on the main board is like due to any of the following process errors: oxygen entered the device and the electrode of the pressure sensor was oxidized and corroded.The wiring inside the pressure sensor was peeled off due to the oxidation corrosion.Foreign matter (epoxy) had adhered to the mounting of the component due to a mistake, the wires inside the pressure-sensor had peeled off due to adherence of the foreign matter.The following is stated in chapter 7.1 of the instructions for use and can detect the event: "how to distinguish abnormal causes and how to cope with them.Details of the error: all leds are off; cause: an error has been detected in the internal circuit of this product." this mdr is being submitted retrospectively as part of a remediation effort related to recent system and process changes.A capa has been opened to manage the actions related to remediation of this issue and any required mdr reporting.
 
Event Description
The customer contacted olympus to report all the lights turned off and the alarm sounded when the cavity mode button was pressed.There was no report of health consequence to a patient.No additional information has been obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11716589
MDR Text Key270921616
Report Number8010047-2021-05312
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/23/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/06/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/05/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-