MAUDE Adverse Event Report: MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO NYLON; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO NYLON; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION Back to Search Results
Model Number NV-3-8-HELIX
Medical Device Problem Codes Separation Failure (2547); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/16/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that the coils had high friction in the middle of the microcatheter, and could not be released.Several detachment attempts were made using the instant and manual detachment methods.The patient's condition necessitating treatment was unknown.It was noted that the patient's blood flow and vessel tortuosity were normal.There were no related patient symptoms.The devices were prepared as indicated per the ifu.There was no damage to the catheter or coil.Ancillary devices include a 6f sheath, 5f cobra guide catheter, direxion 21 microcatheter.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information received reported there was no damage observed to the coil pushwires after removal.
 
Additional Manufacturer Narrative
H3.Product analysis findings: a concerto implant coil was returned within a shipping box and within a sealed tyvek biohazard bag.The concerto implant coil was returned without the introducer sheath.The actuator interface was found to be missing and not attached to the coupler tube.There appeared to be evidence of mechanical detachment using an instant detacher at this location.The concerto pushwire was found to be broken at break indicator, bent at ~19.0cm and at ~134.3cm from broken proximal end.This is indicative that a manual detachment was attempted at these locations.The coin was found still against the lumen stop with what appeared to be blood underneath the outer jacket.The shield coil was found intact with the implant coil still attached.The implant coil was found stretched and damaged with the polypropylene filament intact.The distal outer jacket was removed, and a manual detachment was then attempted by retracting the release wire but was unsuccessful.The coin was removed and found to be broken within lumen stop.The lumen stop was found to be damaged and unable to be measured.The retainer ring was found to be damaged, and the inner diameter was unable to be measured.The detachment stick appeared to have saline residue.No other anomalies were observed.Based on the analysis performed, the concerto coil was confirmed to have non-detachment as the pushwire was returned with the implant coil still attached.However, the implant coil was successfully detached manually during testing.The likely cause for the non-detachment is the coin becoming stuck within the lumen stop.Other possible causes for the non-detachment are the bends and kinks found on the pushers, causing the release wire to become stuck.Additional possible causes for coil and pusher damage are patient vessel tortuosity, device was not hydrated, continuous flush was not administered during procedure, advancing device against resistance or damaged during shipment as the device was returned without its dispenser coil.Since the instant detacher was not returned, any contributing factors of the instant detacher to the non-detachment could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONCERTO NYLON
Common Device Name
DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
Manufacturer (Section D)
MICRO THERAPEUTICS, INC. DBA EV3
9775 toledo way
irvine CA 92618
MDR Report Key11716591
Report Number2029214-2021-00474
Device Sequence Number3591791
Product Code KRD
UDI-Device Identifier00847536033528
UDI-Public00847536033528
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K090046
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/16/2024
Device Model NumberNV-3-8-HELIX
Device Catalogue NumberNV-3-8-HELIX
Device Lot NumberB162093
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2021
Supplement Date Received by Manufacturer04/25/2021
07/14/2021
Initial Report FDA Received Date04/23/2021
Supplement Report FDA Received Date05/12/2021
07/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-