MICRO THERAPEUTICS, INC. DBA EV3 CONCERTO NYLON; DEVICE, VASCULAR, FOR PROMOTING EMBOLIZATION
|
Back to Search Results |
|
| Model Number |
NV-3-8-HELIX |
| Medical Device Problem Codes |
Separation Failure (2547); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received a report that the coils had high friction in the middle of the microcatheter, and could not be released.Several detachment attempts were made using the instant and manual detachment methods.The patient's condition necessitating treatment was unknown.It was noted that the patient's blood flow and vessel tortuosity were normal.There were no related patient symptoms.The devices were prepared as indicated per the ifu.There was no damage to the catheter or coil.Ancillary devices include a 6f sheath, 5f cobra guide catheter, direxion 21 microcatheter.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information received reported there was no damage observed to the coil pushwires after removal.
|
| |
|
Additional Manufacturer Narrative
|
|
H3.Product analysis findings: a concerto implant coil was returned within a shipping box and within a sealed tyvek biohazard bag.The concerto implant coil was returned without the introducer sheath.The actuator interface was found to be intact and attached to the coupler tube.There appeared to be no evidence of mechanical detachment using an instant detacher at this location.The concerto pushwire was found to be kinked at ~18.8cm from proximal end.This is indicative that a manual detachment was attempted at this location.The coin was found against the lumen stop.The shield coil was found intact with the implant coil still attached.The implant coil was found stretched and damaged with the polypropylene filament intact.The distal outer jacket was removed, and a manual detachment was then attempted by retracting the release and was successful.No other anomalies were observed.The lumen stop inner diameter (id) was measured to be 0.0056¿ and found to be within specification (0.0056" minimum - 0.0074" maximum).The retainer ring inner diameter was measured to be 0.0050¿ and found to be within specification.No other anomalies were observed.Based on the analysis performed, the concerto coil was confirmed to have non-detachment as the pushwire was returned with the implant coil still attached.However, the implant coil was successfully detached manually during testing.Likely causes for the non-detachment are the bends and kinks found on the pusher, causing the release wire to become stuck.Additional possible causes for coil and pusher damage are patient vessel tortuosity, device was not hydrated, continuous flush was not administered during procedure, advancing device against resistance or damaged during shipment as the device was returned without its dispenser coil.Since the instant detacher was not returned, any contributing factors of the instant detacher to the non-detachment could not be determined.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|