• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTHXSFT0406
Medical Device Problem Codes Separation Failure (2547); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2021-00837, 3005168196-2021-00839, 3005168196-2021-00840, 3005168196-2021-00841, 3005168196-2021-00842.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the internal carotid artery (ica) using penumbra smart coils (smart coils), penumbra smart coil detachment handles (handles), and two non-penumbra microcatheters.During the procedure, the first smart coil would not detach after multiple attempts were made with three handles.While retracting the smart coil, the physician experienced resistance, and the smart coil was stuck in the first microcatheter.The smart coil was then removed.Subsequently, the physician used a second smart coil, the third handle, and the same microcatheter.However, the physician was also unable to detach the smart coil with the handle after multiple attempts were made.While retracting the smart coil, the physician experienced resistance, and the smart coil was stuck in the microcatheter.The smart coil and microcatheter were then removed.The physician resumed the procedure and successfully placed the third smart coil with the handle and second microcatheter.Next, the physician used a fourth smart coil, the same handle, and the same second microcatheter.However, the physician was also unable to detach the smart coil with the handle after multiple attempts were made and removed the smart coil.It was reported that this was the last attempted embolization.Therefore, the procedure ended at this point.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11716632
Report Number3005168196-2021-00838
Device Sequence Number19840217
Product Code HCG
UDI-Device Identifier00814548015811
UDI-Public00814548015811
Combination Product (Y/N)Y
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/25/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/09/2023
Device Model Number400SMTHXSFT0406
Device Catalogue Number400SMTHXSFT0406
Device Lot NumberF84552
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/25/2021
Initial Report FDA Received Date04/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-