MAUDE Adverse Event Report: CARDINAL HEALTH NPWT FACILITY NPWT SVED THERAPY DEVICE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH NPWT FACILITY NPWT SVED THERAPY DEVICE; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 6701132
Medical Device Problem Codes Melted (1385); Overheating of Device (1437)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation into this reported incident is on-going while we continue to gather further details related to the event.A follow-up report will be filed once the results have been completed.
 
Event or Problem Description
The distributor reported the pcb board and wires are burned and the device needs to be serviced.No other information was provided.
 
Additional Manufacturer Narrative
Supplemental report being filed since sample device received for evaluation.Device released from qa to stock on (b)(6) 2019.No service reports were available at the time of this review.Job router was reviewed without issue.Device passed all tests.Complaint confirmed.Upon initial cosmetic defect inspection damage identified on the device case.When interior of device opened up for evaluation the pcb and wires were burned.Root cause is pump failure.Capa-ma029-01812 was opened and an investigation into the overheating complaints was conducted.The following corrective actions were identified to address the root causes identified in the capa: these corrective actions are planned to be completed by (b)(6) 2021.Ifu for sved device to be updated to include the warning for overheating and no direct contact of the device with patient.Update the service procedure to consider changing the motor is the motor has serviced more than 2000 hours.Implement the service timer recording in the service procedure and forms to determine when to change the motor once the serviceable life has been exhausted.Install relay cut off mechanism at appropriate threshold temperature that will cut off the current to the motor in cases when motor may draw excessive current leading to overheating effect.Cardinal health will continue to monitor the trend of this type of incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NPWT SVED THERAPY DEVICE
Common Device Name
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
CARDINAL HEALTH NPWT FACILITY
8322 eagle palm drive
riverview FL
MDR Report Key11720363
Report Number1423537-2021-00624
Device Sequence Number17110991
Product Code OMP
UDI-Device Identifier10885380097454
UDI-Public10885380097454
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number6701132
Device Catalogue Number6701132
Device Lot Number22381
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/29/2021
Supplement Date Received by Manufacturer03/29/2021
Initial Report FDA Received Date04/23/2021
Supplement Report FDA Received Date05/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-