U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VENITI, INC. VICI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VENITI, INC. VICI Back to Search Results
Model Number 26930
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/11/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a stent migration occurred, requiring new stent placement.Venogram and ivus imaging were performed on an acute deep vein thrombosis (dvt), prior to a stenting procedure.The inferior vena cava (ivc) measured 25mm and the left iliac measured 15.6-18mm.Prior to stenting, thrombectomy and ballooning were performed on the lesion.A 16 x 60mm x 100cm vici self-expanding stent was selected for treatment and was placed in the left iliac/ivc confluence.Upon deployment, the stent fully expanded and apposed the vein walls.The stent was post-dilated.There were no difficulties during the implant procedure and the patient was discharged.The patient returned for a follow-up appointment nine days later, with dvt.It was found that the stent migrated distally, about 4cm.The vici stent was successfully re-stented with a new 18 x 90 wallstent on the right and a 20 x 80 wallstent on the left.The patient is stable and no additional adverse patient effects were reported.Boston scientific (bsc) was only made aware of this event from the physician on april 12, 2021.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VICI
Manufacturer (Section D)
VENITI, INC.
4025 clipper court
fremont CA 94538
MDR Report Key11720489
Report Number2134265-2021-05180
Device Sequence Number17102096
Product Code QAN
UDI-Device Identifier00852725008157
UDI-Public00852725008157
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
P180013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 04/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Model Number26930
Device Catalogue Number26930
Device Lot Number0019100036
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/12/2021
Initial Report FDA Received Date04/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
92672766-FA
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age37 YR
-
-