| Model Number |
22 GAUGE |
| Medical Device Problem Code |
Separation Failure (2547)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
03/18/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Information received a smiths medical implantable ports|deltec gripper needle separated and embedded under the skin inside port.The nurse went to access portacath and noticed child was scratching site, swab obtained and redressed.On flushing the portacath it was noticed it was leaking.Nurse asked another member of staff to review.She tried to replace the port needle, however the needle had not came off with the gripper and was left inside skin.The plastic gripper section was reported to coma away.The patient was then transferred to another hospital to have the needle removed under general anesthetic.Needle was successfully removed.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: event problem and evaluation codes: updated after device evaluation.H10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Units received: sixteen (16) samples were received for evaluation; the returned samples were received decontaminated and inside a plastic fifteen (15) samples received with its original closed package and one (1) sample was received in used condition without its original package.Visual inspection results: it is observed in one sample the needle detached for the base.The complaint is confirmed.Replication of the failure mode: a needle gripper was bent and tested using fixture t98-3198 to proof if the foam assembly could be performed with bent needle condition.Results: the bent samples could not fix in the t98-3198 cavities to perform the foam assembly.The root cause is the product became damaged after it left shm facilities.The cause of the reported problem could not be determined.
|
| |
|
Search Alerts/Recalls
|