| Model Number |
48233540 |
| Medical Device Problem Codes |
Material Separation (1562); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the tulips of 2 xia uniplanar screws disengaged from their respective screw shafts intra-operatively.Subsequently, there was a 60 minute delay to surgery as the surgeon removed the screws and rods on both sides to redo the construct.The procedure was completed successfully.This report captures the first of the two screws.
|
| |
|
Event or Problem Description
|
|
It was reported that the tulips of 2 xia uniplanar screws disengaged from their respective screw shafts intra-operatively.Subsequently, there was a 60 minute delay to surgery as the surgeon removed the screws and rods on both sides to redo the construct.The procedure was completed successfully.This report captures the first of the two screws.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has been received and upon receipt d4 (lot #) has been corrected from '190449' to '197153'.
|
| |
|
Additional Manufacturer Narrative
|
|
B.5 has been corrected following new information received and investigation completion.Visual inspection confirmed that tulip did not disengage and that there is no deformation on this product.Report 0009617544-2021-00067 addresses the complaint of the tulip disengagement, as visual inspection confirmed that the tulip did disengage in that record.
|
| |
|
Event or Problem Description
|
|
It was initially reported that the tulips of 2 xia uniplanar screws disengaged from their respective screw shafts intra-operatively.However, after additional information received and investigation completion, 8 screw tulips disengaged.Subsequently, there was a 60 minute delay to surgery as the surgeon removed the screws and rods on both sides to redo the construct.The procedure was completed successfully.This report captures the first returned screw which shows no signs of tulip disengagement.
|
| |
|