STRYKER SPINE-US XIA DEFORMITY UNIPLANAR SCREW DIAM 5.5 X 40 MM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Model Number 48233540
Medical Device Problem Codes Material Separation (1562); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/26/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the tulips of 2 xia uniplanar screws disengaged from their respective screw shafts intra-operatively.Subsequently, there was a 60 minute delay to surgery as the surgeon removed the screws and rods on both sides to redo the construct.The procedure was completed successfully.This report captures the second of the two screws.
 
Event or Problem Description
It was reported that the tulips of 2 xia uniplanar screws disengaged from their respective screw shafts intra-operatively.Subsequently, there was a 60 minute delay to surgery as the surgeon removed the screws and rods on both sides to redo the construct.The procedure was completed successfully.This report captures the second of the two screws.
 
Additional Manufacturer Narrative
D4 has been updated following receipt of the device.
 
Additional Manufacturer Narrative
B.5 has been corrected following new information received and investigation completion.Visual: visual inspection confirmed tulip disengagement.Deformation was observed on shank bulb from the rod.This deformation is on the outer edge of shank bulb.The direction of deformation also indicates that the rod axis was not in line with the uniplanar axis of the tulip.Additional deformation is seen on the star patterned driver engagement surface.Deformation is observed on the locking ring of the tulip.Deformation on locking ring is observed in more than one position.Device history records were reviewed for this lot, no relevant manufacturing issues were identified.Complaint history records were reviewed for this lot, similar complaints were identified.Visual inspection shows that multiple maneuvers were performed simultaneously or consecutively that put excess stress on the screw tulip.Evidence of this is shown in the multiple deformations on the tulip locking ring as well as the deformation left on the screw shank bulb, which was out of line with the tulip axis.Likely cause is due to excessive angulation and excessive force applied as seen in the deformation on the screw and tulip locking ring.
 
Event or Problem Description
It was initially reported that the tulips of 2 xia uniplanar screws disengaged from their respective screw shafts intra-operatively.However, after additional information received and investigation completion, 8 screw tulips disengaged.Subsequently, there was a 60 minute delay to surgery as the surgeon removed the screws and rods on both sides to redo the construct.The procedure was completed successfully.This report captures the sixth of the eight screws.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA DEFORMITY UNIPLANAR SCREW DIAM 5.5 X 40 MM
Common Device Name
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key11721014
Report Number0009617544-2021-00067
Device Sequence Number17101968
Product Code NKB
UDI-Device Identifier07613327001037
UDI-Public07613327001037
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
K142381
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number48233540
Device Catalogue Number48233540
Device Lot Number197153
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2021
Initial Date Received by Manufacturer 03/26/2021
Supplement Date Received by Manufacturer05/04/2021
06/29/2021
Initial Report FDA Received Date04/23/2021
Supplement Report FDA Received Date05/25/2021
07/12/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-