Model Number SERVO-S
Device Problem Decrease in Pressure (1490)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/30/2021
Event Type  malfunction  
Event Description
It was reported that the ventilator had a low oxygen pressure.The patient involvement is unknown.Manufacturer's ref.#: (b)(4).
 
Manufacturer Narrative
The ventilator was investigated on site and the o2 gas module was replaced and returned for further investigation.No device log file has been received.The reported failure with an alarm for low inlet oxygen pressure was reproduced during simulated use testing of the returned oxygen gas module in a ventilator.A defective pressure transducer on a printed circuit board inside the gas module caused the failure.The supply pressure transducer is part of the flow measuring in the gas module.The failure of the pressure transducer leads to inaccurate gas flow regulation which will be detected during pre-use check and alarms will be activated if the failure occurs during ventilation.The manufacture date has been wrongly reported in the initial mdr and is corrected accordingly.
 
Event Description
Manufacturer's ref.#: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SERVO-S
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
sara soderholm roman
roentgenvagen 2
solna 
MDR Report Key11722828
MDR Text Key247214602
Report Number8010042-2021-00915
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K123149
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSERVO-S
Device Catalogue Number6640440
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received03/01/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/27/2014
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-