According to the reporter, during deposition of vena cava in a laparoscopic adrenalectomy, the clip could be placed, but the clip did not came off the black holder, so that the clip and black holder hung on the vena cava.Surgeon used titanium clips to resolve the issue.No patient injury.
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.Two devices were available for evaluation.Visual inspection noted the clips were fully formed but not completely deployed from the cartridges.It was reported that the jaws of the device remained closed on tissue after firing.The reported issue was confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if the trigger of the instrument is not fully actuated releasing the clip from the cartridge.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.The instructions included with this device provide the following guidance: reusable devise must be carefully inspected prior to and after each use, as any damage may affect the safe operation of the device.Failure to lubricate the device prior to sterilization may cause malfunctions such as jamming, difficult operation, or unacceptable staple formation.Failure to fully squeeze the handle may result in miss-formed clips which may result in incomplete closure and lack of hemostasis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|