| Model Number |
MRDP66PLS |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
01/30/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the sterilization certificates for the loads the lots were processed on indicated that the loads passed all requirements of sterilization.There were no non-conformances during the sterilization process.Sterilization of medcomp product is performed according to iso standard 11135 "medical devices- validation and routine control of ethylene oxide sterilization".Infections are the most common complication after implantation of a venous port system.Infections of port venous systems include pocket and/or tunnel cellulitis or the more common catheter-related blood stream infections.Instructions for use contain the following potential complication- catheter or port related sepsis.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
The facility submitted a voluntary event report stating there were three bloodstream infections in oncology patients with port over 18 day period.No other information was provided.
|
| |
|