MAUDE Adverse Event Report: MEDICAL COMPONENTS, INC. 6.6F DIGNITY LP CT PORT; DIGNITY CT-PORT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDICAL COMPONENTS, INC. 6.6F DIGNITY LP CT PORT; DIGNITY CT-PORT Back to Search Results
Model Number MRDP66PLS
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 01/30/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A review of the sterilization certificates for the loads the lots were processed on indicated that the loads passed all requirements of sterilization.There were no non-conformances during the sterilization process.Sterilization of medcomp product is performed according to iso standard 11135 "medical devices- validation and routine control of ethylene oxide sterilization".Infections are the most common complication after implantation of a venous port system.Infections of port venous systems include pocket and/or tunnel cellulitis or the more common catheter-related blood stream infections.Instructions for use contain the following potential complication- catheter or port related sepsis.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
The facility submitted a voluntary event report stating there were three bloodstream infections in oncology patients with port over 18 day period.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.6F DIGNITY LP CT PORT
Common Device Name
DIGNITY CT-PORT
Manufacturer (Section D)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
MDR Report Key11723627
Report Number2518902-2021-00023
Device Sequence Number3894625
Product Code LJT
UDI-Device Identifier00884908149890
UDI-Public884908149890
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2024
Device Model NumberMRDP66PLS
Device Catalogue NumberMRDP66PLS
Device Lot NumberMNXJ480, MPFN020
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2021
Initial Report FDA Received Date04/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-