| Model Number |
MX5342L |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/28/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that while the medex stopcock was in use with a patient, the patient's blood pressure dropped significantly from a map's of 70-80's down to 50's.The operator discovered that the pressor support medication (norepinephrine) was leaking out of the cracked stop cock.The operator did not have to give epinephrine as a result.The issue was resolved once the operator found the source of the damaged stop cock and moved the norepinephrine.No further complications were reported in relation to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Returned device was received for evaluation.During the evaluation of the device the customer reported condition was confirmed, root cause could not be attributable to manufacturing site, the most likely root cause of the condition observed on returned samples indicate a possible mishandling of during use, as force is required to break the devices.Problem source was traced to user interface.Lot number was not provided therefore dhr review cannot be executed.
|
| |
|