| Model Number |
7211 |
| Medical Device Problem Codes |
Deflation Problem (1149); Entrapment of Device (1212); Difficult to Remove (1528); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/05/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that catheter entrapment and removal difficulties were encountered.The 80% stenosed target lesion was located in the non-tortuous and moderately calcified left main and proximal left anterior descending artery.The physician implanted a 4.0mm stent and a 5.00mm x 8mm nc emerge balloon catheter was used for post-dilatation.However, during deflation of balloon, resistance was felt in removing the balloon from the stent and multiple attempts were made to deflate the balloon to its smallest profile.The tip and proximal of the balloon was kinked and the balloon became stuck on the guidewire.After full deflation, the physician pulled the balloon very hard and the balloon was removed from the left main stent.The balloon and guidewire were removed together intact and the balloon catheter was noted to be shortened or wrinkled.The procedure was completed with another of same device.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that catheter entrapment and removal difficulties were encountered.The 80% stenosed target lesion was located in the non-tortuous and moderately calcified left main and proximal left anterior descending artery.The physician implanted a 4.0mm stent and a 5.00mm x 8mm nc emerge balloon catheter was used for post-dilatation.However, during deflation of balloon, resistance was felt in removing the balloon from the stent and mulitple attempts were made to deflate the balloon to its smallest profile.The tip and proximal of the balloon was kinked and the balloon became stuck on the guidewire.After full deflation, the physician pulled the balloon very hard and the balloon was removed from the left main stent.The balloon and guidewire were removed together intact and the balloon catheter was noted to be shortened or wrinkled.The procedure was completed with another of same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of an nc emerge balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was a complete separation at 48.9cm distal of the strain relief.There was no proximal portion of the device.There was contrast and blood in the inflation lumen and contrast in the balloon, and blood in the guidewire lumen.The balloon was loosely folded.The inner shaft was buckled 50mm from the tip.The outer shaft was severely stretched down.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis confirmed the reported difficulty removing the device.
|
| |
|