| Model Number |
429688 |
| Medical Device Problem Code |
Unstable Capture Threshold (3269)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: 457445 lead implanted: (b)(6) 2004.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the left ventricular (lv) lead exhibited variable thresholds.It was also reported that the right ventricular (rv) lead exhibited unstable thresholds and was not reliable and might be moving.The patient was noted to have passed out during testing of the cardiac resynchronization therapy pacemaker (crt-p) feature which monitors pacing thresholds at periodic intervals.It was noted that only ventricular support paces are delivered during testing versus bi-ventricular pacing support.The feature was turned off with no further symptoms.The lv lead, rv lead and crt-p remain in use.No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated pacing capture threshold in the left ventricle was high.The device memory indicated pacing capture threshold in the left ventricle was unstable.Correction: b2 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additionally, it was reported that reprogramming was performed for the leads.It was further clarified that the rv lead moving was in relationship to moving/variable thresholds measurement.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: h3; h6 eval code result (fdr/annex c), eval code result (fdr/annex d).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|