• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PERIPHERAL ROTAWIRE AND WIRECLIP TORQUER; CATHETER, PERIPHERAL, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PERIPHERAL ROTAWIRE AND WIRECLIP TORQUER; CATHETER, PERIPHERAL, ATHERECTOMY Back to Search Results
Model Number 29710
Medical Device Problem Code Entrapment of Device (1212)
Health Effect - Clinical Codes Myocardial Infarction (1969); Vasoconstriction (2126)
Date of Event 04/05/2021
Type of Reportable Event Death
Event or Problem Description
It was reported that the rotaburr became stuck within the lesion, the rotaburr and rotawire were difficult to remove, the procedure was aborted, the patient went to surgery then later died during surgery.A 2.00mm peripheral rotalink plus atherectomy catheter burr and a peripheral rotawire guidewire were selected for use in a tibial atherectomy procedure.The patient was being treated at an outpatient facility.The 100% stenosed target lesion was located in the mildly tortuous and severely calcified peroneal artery.The devices were advanced to the lesion and atherectomy was performed with a set rotational speed of 180,000rpm.The burr stalled and became lodged into the peroneal artery.The burr could not be removed from the peroneal artery.At the time the burr became stuck in the peroneal artery the patient experienced a vasospasm which may have played a role in the burr becoming stuck within the peroneal artery.The procedure was cancelled and the patient was transferred via an ambulance to the hospital to have the devices removed surgically.The peripheral rotawire guidewire was removed but the patient experienced a cardiac event including myocardial infarction.During the burr removal, the patient died.The burr was not recovered.There was no postmortem performed on the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERIPHERAL ROTAWIRE AND WIRECLIP TORQUER
Common Device Name
CATHETER, PERIPHERAL, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11726882
Report Number2134265-2021-05021
Device Sequence Number3946079
Product Code MCW
UDI-Device Identifier08714729863649
UDI-Public08714729863649
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/19/2023
Device Model Number29710
Device Catalogue Number29710
Device Lot Number0026659023
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/05/2021
Initial Report FDA Received Date04/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Other; Required Intervention;
Patient Age70 YR
Patient Weight64
-
-