Model Number 66022006
Medical Device Problem Code Pumping Stopped (1503)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, after treatment with four pico 7 15cmx20cm the pump stopped working within an hour.It was unknown how the procedure was finished.No patient injuries or other complications were reported.
 
Additional Manufacturer Narrative
The information received by the manufacturer has been re-evaluated for mdr reporting and it was determined that this case does not meet the threshold for reporting and is a non-reportable event.The loss of negative pressure wound therapy due to an inoperative or 'stopped' pump will not directly cause or contribute to serious injury or death as the pico dressing can revert to managing the wound as a standard multilayer dressing.No risk to patient safety is anticipated.This event is considered not reportable pursuant to 21 cfr part 803.If further details are provided confirming the occurrence of a reportable event, our files will be updated accordingly and a further report submitted outlining both the event details and our investigations performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PICO 7 15CMX20CM
Common Device Name
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11726890
Report Number8043484-2021-00939
Device Sequence Number4026232
Product Code OMP
UDI-Device Identifier05000223492212
UDI-Public05000223492212
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K111170
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number66022006
Device Catalogue Number66022006
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/04/2021
Supplement Date Received by Manufacturer08/06/2021
Initial Report FDA Received Date04/26/2021
Supplement Report FDA Received Date08/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-