| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/31/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Customer has not indicated whether the product will be returned or not, to zimmer biomet for investigation.Additional requests have been sent to the customer.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported the cps poly broke.Attempts have been made and no further information has been provided at this time.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Complaint sample was evaluated.Visual evaluation of the returned device shows signs of repeated use and is fractured on medial side of post.Device history record (dhr) was reviewed and no discrepancies relevant to the reported event were found.Evaluation of the returned device identified the fracture was consistent with the tasp fractures analyzed in previous investigation.It identified that the common failure modes for the tasp devices include either bending overload or low cycle fatigue culminating in bending overload as evident by the presence of hackle marks, river lines and striations features on the fracture surface.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|