• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO HANDPIECE; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number PFSR110137
Device Problems Failure to Calibrate (2440); Naturally Worn (2988); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2021
Event Type  malfunction  
Event Description
It was reported that, during set up of a navio ukr procedure, the handpiece failed to calibrate multiple times.They switched pointer probe and tracker array, and eventually switched handpiece.After switching the handpiece, everything calibrated fine.The groves on the handpiece are severely worn (damaged clamshell) and it seems like the pointer probe is not seating properly.Since incident occurred before the procedure, there was no patient involvement.No other complications were reported.
 
Manufacturer Narrative
The navio handpiece part number 110137, sn (b)(6) intended for treatment was returned for evaluation.A relationship between the reported event and the device was confirmed.The reported problem was visually confirmed.The point probe grooves are very worn.Although the reported problem was visually confirmed, a functional evaluation was performed.The handpiece passed all tests.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.The most likely cause of this event is wear and tear of the point probe grooves over time.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO HANDPIECE
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key11728623
MDR Text Key248363745
Report Number3010266064-2021-00301
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556628515
UDI-Public00885556628515
Combination Product (y/n)N
PMA/PMN Number
K152574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPFSR110137
Device Catalogue NumberPFSR110137
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2021
Date Manufacturer Received08/03/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-