• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number 7N8391
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/25/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device manufacturer address: (b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that a one-link catheter extension set was leaking from the junction between the hub and the line extension.The leak was discovered while the patient was connected for contrast injection.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
H10: the actual device was not available; however, a companion sample was received for evaluation.Visual inspection using the naked eye did not identify any abnormalities that could have contributed to the reported condition.All components were correctly placed and according to specifications.The sample was connected at the female and male luer connector and functional testing was performed including clear passage and pressure testing; and the device performed according to product specifications.The reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTENSION SETS WITH ONE-LINK NEEDLE-FREE LV CONNECTOR
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key11729177
Report Number1416980-2021-02436
Device Sequence Number11517738
Product Code FPA
UDI-Device Identifier00085412475400
UDI-Public(01)00085412475400
Combination Product (Y/N)Y
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K132734
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/26/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue Number7N8391
Device Lot NumberR20L08088
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/30/2021
Supplement Date Received by Manufacturer05/25/2021
Initial Report FDA Received Date04/26/2021
Supplement Report FDA Received Date05/26/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-