| Model Number |
21-7624-24 |
| Medical Device Problem Code |
Audible Prompt/Feedback Problem (4020)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/26/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Customer did not provide the lot number of the affected device.
|
| |
|
Event or Problem Description
|
|
Information was received regarding a cadd administration set.It was reported that during the use of the product, an occlusion alarm sounded.After the user changed the cassette to another one, the alarm was settled.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Other, other text: h6: event problem and evaluation codes: updated after device evaluation h10: device evaluation: the device was returned for investigation.A visual inspection and functional test were performed.Visual inspection: no occlusions nor obstructions were detected on the samples.Functional test: the complaints sample was tested using the cadd pump solis [cal.Id: 1.0326; due date: august, 2021] to prime the devices as instruction of use as10003589-004 rev.100 indicate.No difficulty were detected during the priming and no alarms were activated, the water could pass through the whole devices; thus the failure mode reported is not confirmed.The cause of the reported problem could not be determined.No corrective actions are required since the complaint was not confirmed.
|
| |
|