| Model Number |
52-3192 |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Pain (1994); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
07/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Pinnacle unf and medical records received.After review of the medical records the patient was revised due to malfunction of mom left hip arthroplasty resulting to pain, elevated cobalt and chromium level and mechanical sensation left tha.Operative note reported milky yellow fluid collection consistent with adverse local tissue reaction with dark staining of the capsular tissues and synovium.Noticeable gross wear of the superior aspect of the acetabular liner.Significant corrosion noted at the locking mechanism of the liner and the shell.However, minimal corrosion noted at the head neck taper junction.Doi: (b)(6) 2008; dor: (b)(6) 2019; left hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).Investigation summary: no device associated with this report was received for examination.The device manufacturing records have been reviewed and no related deviations or anomalies were identified.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot: product code 523192 work order (b)(4) was manufactured on 12-apr-2008.(b)(4) parts were manufactured per specification and all raw materials met specification.There were no nrs (non-conformance records) associated with lot 2584820.Device history review : product code 523192 work order (b)(4) was manufactured on 12-apr-2008.(b)(4) parts were manufactured per specification and all raw materials met specification.There were no nrs (non-conformance records) associated with lot 2584820.
|
| |
|
Search Alerts/Recalls
|