Model Number 1054655
Device Problem Decrease in Pressure (1490)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/07/2021
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator was alarming for a "service required" condition related to low inspiratory pressure.There was no harm or injury reported.The ventilator was returned to the manufacturer for evaluation, and the customer's complaint was confirmed.The device's overhead blower filter and active exhalation control module need replaced to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 100
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS INC
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS INC
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
adam price
312 alvin drive
new kensington, PA 15068
7243349303
MDR Report Key11732374
MDR Text Key247538969
Report Number2518422-2021-01216
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00606959025387
UDI-Public00606959025387
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K083526
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 04/27/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1054655
Device Catalogue Number1054655
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/07/2021
Date Manufacturer Received04/07/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-