| Model Number |
391020 |
| Medical Device Problem Codes |
Failure to Advance (2524); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Event or Problem Description
|
|
An orsiro drug-eluting stent system was selected for treatment.The device could not cross the mild calcified lesion (90 percent stenosis degree) in the severe tortuous distal lad and the stent was deformed.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.The returned device was subjected to a detailed technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for this event.The technical investigation revealed that stent and balloon show no irregularities.A stent deformation could not be confirmed, the crimped stent diameter is in specification.The review of the production documentation of the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.As a part of final inspection every stent system undergoes visual inspection to ensure correct embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.It should be noted that the ifu clearly states that a predilatation of the target lesion is mandatory.
|
| |
|