| Model Number |
GN060 |
| Medical Device Problem Code |
Power Problem (3010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a capa is not necessary.
|
| |
|
Event or Problem Description
|
|
It was reported to aesculap ag that a bipolar coagulation unit (part # gn060) was tested prior to a procedure set to be performed on (b)(6) 2021.According to the complainant, prior to the operation, the bipolar coagulator was tested, and was found to have no output.The device was replaced.The complaint device was not returned to the manufacturer for evaluation.No patient involvement.Additional information was not provided nor available / was not available.The malfunction is filed under aag reference (b)(4).
|
| |
|
Search Alerts/Recalls
|