| Model Number |
00884450320563 |
| Medical Device Problem Codes |
Material Fragmentation (1261); Migration (4003)
|
| Health Effect - Clinical Codes |
Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The suspect device is not expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
|
| |
|
Event or Problem Description
|
|
The account alleges that during a hemodialysis reliable outflow graft de-clot procedure on (b)(6) 2021 , the physician noted that a distal part of the voc placed within this patient in 2017 had detached.According to radiology images received from the account, approximately 5cm of the distal voc is free floating within the patient's right ventricle.Intervention will be required for foreign body removal.Patient was admitted and asymptomatic.
|
| |
|
Additional Manufacturer Narrative
|
|
The suspect device was not returned for evaluation.A photograph was provided by the account but the location of the break or detachment is unclear.The complaint could not be confirmed.The root cause could not be determined.A review of the device history and complaint database could not be performed since the lot number was not provided.
|
| |
|