Model Number 71674084
Device Problem Material Twisted/Bent (2981)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/06/2021
Event Type  malfunction  
Event Description
It was reported that during an internal fixation surgery (retrograde nail), it was noticed that the tip of a screwdriver release handle is completely bent.No patient injury was reported.There was no significant delay and the procedure was finished using a smith & nephew back up.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned for evaluation.A visual inspection confirmed the tip of the screwdriver release handle appears to be bent.This device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the complaint history on the listed part revealed no prior complaints for the listed failure mode with the same batch number.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A relationship between the device and the reported incident was corroborated.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREWDRIVER RELEASE HANDLE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11734983
MDR Text Key247682165
Report Number1020279-2021-03541
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010560810
UDI-Public03596010560810
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/27/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71674084
Device Catalogue Number71674084
Device Lot Number18JM09047
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-