• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN DISTAL FEM CUTTING BLOCK; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN DISTAL FEM CUTTING BLOCK; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED. Back to Search Results
Model Number 71434403
Medical Device Problem Codes Naturally Worn (2988); Incomplete or Inadequate Connection (4037)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/06/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a tka the lgn distal fem cutting block was worn out and did not stay tightened down.The procedure was successfully completed without delay using the same device.This happened when the instruments were out of the patient hence no patient injury or other complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The hex screw slot of the device's knob is severely worn preventing the device from functioning as intended.The device also has several nicks and scratches as well as wear.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN DISTAL FEM CUTTING BLOCK
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED.
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11735232
Report Number1020279-2021-03546
Device Sequence Number17150446
Product Code HRY
UDI-Device Identifier03596010550163
UDI-Public03596010550163
Combination Product (Y/N)N
Initial Reporter StateMT
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71434403
Device Catalogue Number71434403
Device Lot Number12CM00399
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/19/2021
Initial Date Received by Manufacturer 04/06/2021
Supplement Date Received by Manufacturer05/18/2021
Initial Report FDA Received Date04/27/2021
Supplement Report FDA Received Date05/18/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-